140
Maintenance
Batteries are nearly empty.
Replace batteries now.
Batteries Empty [M-31]
19 Jul 2018 10:10 AM
Maintenance
Meter too cold or too warm.
Power meter off. Repeat test
after 30 minutes at room
temperature.
Ambient Temperature
[M-32]
19 Jul 2018 10:10 AM
Maintenance
No blood was detected.
Repeat test, with a new test
strip if necessary.
No Blood Detected
[M-41]
19 Jul 2018 10:10 AM
Maintenance
Not enough sample was
applied. Repeat test with a
new test strip.
Sample Too Small
[M-42]
19 Jul 2018 10:10 AM
Maintenance
A dosing error was detected.
Repeat test with a new test
strip.
Dosing Error [M-43]
19 Jul 2018 10:10 AM
Maintenance
Test strip failed internal
control. Repeat test with a
new test strip.
Test Strip Error [M-44]
19 Jul 2018 10:10 AM
Maintenance
A USB error has occurred.
Power meter off and on and
repeat connecting USB.
USB Error [M-71]
19 Jul 2018 10:10 AM
Maintenance
Pairing has failed. Repeat
pairing process.
Bluetooth Error [M-72]
19 Jul 2018 10:10 AM
Symbols, Icons, Messages
CoaguChek Vantus User’s Manual, version 1.0