Remote Control - Commands R&S NRP-Z11/21/31/41/61/211/221
1137.7470.12 6.32 E-7
Command Parameter Unit Default setting Page
SENSe:CORRection:OFFSet:STATe[?] OFF | ON OFF 6.10
SENSe:CORRection:SPDevice:STATe[?] OFF | ON OFF (can be
modified by the user)
6.10
SENSe:FREQuency[?] 10.0e6 to 8.0e9 (NRP-Z11/211) or
10.0e6 to 18.0e9 (NRP-Z21/221) or
10.0e6 to 33.0e9 (NRP-Z31) or
10.0e6 to 40.0e9 (NRP-Z41) or
10.0e6 to 50.0e9 (NRP-Z61)
Hz 50.0e6 6.11
SENSe:FUNCtion[?] "POWer:AVG" |
"POWer:TSLot:AVG" |
"POWer:BURSt:AVG" |
"XTIMe:POWer"
"POWer:AVG" 6.11
SENSe:POWer:AVG:APERture[?] 10.0e–6 to 0.3 s 0.02 6.12
SENSe:POWer:AVG:BUFFer:SIZE[?] 1 to 1024 1 6.13
SENSe:POWer:AVG:BUFFer:STATe[?] OFF | ON OFF 6.13
SENSe:POWer:AVG:SMOothing:STATe[?] OFF | ON OFF 6.13
SENSe:POWer:BURSt:DTOLerance[?] 0.0 to 0.003 s 0.0001 6.14
SENSe:POWer:TSLot:AVG:COUNt[?] 1 to 128 8 6.14
SENSe:POWer:TSLot:AVG:WIDTh[?] 10.0e–6 to 0.1 s 0.001 6.14
SENSe:RANGe[?] 0 to 2 (NRP-Z11/21/31/41/61) or
0 to 1 (NRP-Z211/221)
2 (NRP-Z11/21/31/
41/61)
1 (NRP-Z211/221)
6.15
SENSe:RANGe:AUTO[?] OFF | ON ON 6.15
SENSe:RANGe:CLEVel[?] –20.0 to 0.0 dB 0.0 6.15
SENSe:SAMPLING[?] FREQ1 | FREQ2 FREQ1 6.15
SENSe:SGAMma:CORRection:STATe[?] OFF | ON OFF 6.16
SENSe:SGAMma:MAGNitude[?] 0.0 to 1.0 0.0 6.17
SENSe:SGAMma:PHASe[?] –360.0 to 360.0 degree 0.0 6.17
SENSe:TIMing:EXCLude:STARt[?] 0.0 to 0.1 s 0.0 6.17
SENSe:TIMing:EXCLude:STOP[?] 0.0 to 0.003 s 0.0 6.17
SENSe:TRACe:AVERage:COUNt[?] 1 to 8192 4 6.17
SENSe:TRACe:AVERage:STATe[?] OFF | ON ON 6.19
SENSe:TRACe:AVERage:TCONtrol[?] MOVing | REPeat REPeat 6.19
SENSe:TRACe:MPWidth? s 6.19