Instrument Construction and Function Description R&S ZVB
1145.1078.82 3.70 E-4
Testing the frequency reference board
(see Test Equipment)
Remove the board from the instrument.
Reinsert board and extension card.
Connect the output to be tested to the spectrum analyzer using an adapter cable and
adapter SMA-N.
Set the frequencies listed in the table.
Check signals according to following table.
Signal Connector Frequency Nom. level Setting
NA1_AD_CLK X103 80 MHz 10 dBm ±1dB
NA2_AD_CLK X104 80 MHz 10 dBm ±1dB
SY1_REF X105 128 MHz 10 dBm ±1dB
SY2_REF X106 128 MHz 10 dBm ±1dB
REF_10_OUTIN X107 10 MHz 6 dBm ±1dB Menu/System/Internal
Reference
The board must be replaced if the signal is more than 2 dB below the stated nominal level.
Connect the frequency reference output of the spectrum analyzer to X107 (REF_10_OUTIN) using
an adapter cable and adapter SMA-N.
Connect X105 or X106 (SY1_REF or. SY2_REF) to the spectrum analyzer using an adapter cable
and adapter R&S SMA-N.
The 128 MHz signal’s frequency must precisely equal its nominal frequency.
If the frequency differs in any way (e.g. 128.001 MHz), replace the board.