Performance Test R&S
ZVB
1145.1078.82 1.6 E–4
Checking the Maximum Output Power
Instrument: Power sensor or
Power meter with power sensor
(see Chapter “Test Equipment“, Item2)
Test setup: Connect power sensor to port1 (port 2 to port 4)
Power sensor settings: For measurement frequencies, see Performance Test Report {f
GEN
*}.
R&S
ZVB settings:
[ Preset ]
[ Meas : Wave Quantities: a1 Src Port 1
(a2 Src Port 2 to a4 Src Port 4)]
[ Sweep : Sweep Type : CW Mode :
CW Frequency : {f
GEN
*} ; Power : 16 dBm]
[ Sweep : Single : Restart]
* For measurement frequencies see the Performance Test Report {f
G
EN
}.
Measurement: Read off the level indicated by the power meter.
Max. power ................................................... see Performance Test Report