EasyManua.ls Logo

Roland E-50 - Touch Screen Calibration

Roland E-50
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-50
25
1.2 Touch Screen Calibration
c
Enter the “INTERNAL TEST” mode (cf. paragraph 1.1)
d
Wait for the initialization of the E-50 to start.
e
Using the touch screen pen, touch for a second, the points shown on the display, the sequenze is the following:
down on the left
up on the right
down on the right
up on the left
f
If the operation is not successful, the display shows NG:Calibration, in that case repeat the operation from scratch (point 3): after the error
message wait until the calibration video appears again
g
If the operation is successful, the display will show Calibration OK and after a while the instrument will pass to the next test.
4 2
d
e
g
1