Quick Start Reference – Calibrating Density
(For Completion of a Single Point Reference)
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Main Menu
Variables
Status Display
Configuration
Digital Outputs
Digital Inputs
Calibration .
Dia
g
nostics
Step 3 ‘Low Reference Density’ Calibration
(Have the process ready filled with water)
Scroll down to “Calibration” Press to enter menu
Scroll down to “Cal Density” Press the to enter menu
Scroll down to “Low Reference” Press the to enter menu
Select “Perform Reference” press to enter menu
Read the next screen carefully and acknowledge by pressing the F4
(OK) Key. Referencing will begin. (This may take several minutes.)
Scroll down to “Ref Density” Press to enter menu
Notice the value that is displayed.
Use Key and Key to toggle back and forth between digits.
Change the value using or
Enter the value of the process (Example 0.0000 %Solids for water).
Press The F4 Key to enter and store the value.
Press the F3 Key (Home) to return to the Main Menu.
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow
Gauge
Low Reference
Perform
Reference
.
Ref Density 0.0000
Ref Cap 20000
HOME
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Calibration
State Fully Calibra
Ref Constants
Cal
Density .
Cal Flow Rate
Temp Reference
Cal Remote Solids
Last Ref.Date
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Cal Density
State
Low Reference
High Calibrate
Manual Entry
Clear Ref/Cal
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Low Reference
WARNING – Remove
loop from automatic
control, verify that
source is on, and the
vessel/pipe is empty.
ABORT OK
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Low Reference
Reference
R
ef
Density 0.0000
Ref Cap 18754
HOME
C
F1 F2 F3 F4
Ronan Engineering
X96S Mass Flow Gauge
Ref Density
0.0000 SGU
0.0000
ESC ENTER