August 2012
18
Quick Start Guide
European approvals
ATEX approvals
E1 Certificate: Sira 05ATEX1129X
Flameproof and Dust:
ATEX Marking II 1/2 G D
Ex d IIC T6...T2 Ga/Gb
Ex tb IIIC T85 °C...T265 °C Db
(See “Instructions specific to hazardous area installations (E1 and E7)” on page 23)
I1 Certificate: Sira 05ATEX2130X
Intrinsic Safety and Dust:
ATEX Marking II 1 G D
Ex ia IIC T5...T2 Ga
Ex ia IIIC T85 °C...T265 °C Da
(See “Instructions specific to hazardous area installations (I1 and I7)” on page 25)
Rest of the world approvals
International Electrotechnical Commission (IEC) approvals
E7 Certificate: IECEx SIR 06.0051X
Flameproof and Dust:
Ex d IIC T6...T2 Ga/Gb
Ex tb IIIC T85 °C...T265 °C Db
(See “Instructions specific to hazardous area installations (E1 and E7)” on page 23)
I7 Certificate: IECEx SIR 06.0070X
Intrinsically Safe and Dust:
Ex ia IIC T5...T2 Ga
Ex ia IIIC T85 °C...T265 °C Da
(See “Instructions specific to hazardous area installations (I1 and I7)” on page 25)
Note
A certified isolating amplifier to IEC 60947-5-6 is required for intrinsic safety if the
NAMUR electronics is used in a hazardous area installation
A certified intrinsically safe barrier is required for intrinsic safety if the 8/16 mA
electronics is used in a hazardous area installation