Piranha/Cobia & Quick Check Getting Started
Quick Check
38
7. Measured data are transferred to Excel for each exposure. You can use the small minimize Excel window to
control your measurement procedure while having Excel on the screen. You can change basic meter settings
and the left most button is used to switch between Quick Check and Excel. Data are arriving in a fixed format
(compatible with tyhe format used by Xi and X2 View). The fixed format is defined below.
Definition of the fixed format
There are four different detector combinations that each has its own fixed format:
· Multi-detector (internal detector) with an optional external detector
· Only an external radiation detector (Dose Probe, CTDP, T20, Ion chamber of any type)
· Only the Light detector
· Only a mAs probe
Multi-detector (internal detector)
Tube voltage
unit
Dose
unit
Dose rate
unit
Exp. time
unit
Pulses
unit
Empty
Empty
Pulse rate
unit
Dose/pulse
unit
HVL
unit
Tube mAs
unit
Tube mA
unit
TF
unit
Dose (ext.)