LRF 7047- Integrator Manual
914929_TML_LRF7047_en_Version B
Public: 04.2019
Confidential & Proprietary Safran Vectronix AG – All rights reserved
Page: 56/72
6.6.6 Reset LRF module to default settings >LM,Rd*<CS><CR>
This command resets all LRF module’s configuration parameters to their default values.
An automatic restart of the LRF module is performed during the process.
Default parameters are:
Baud rate 57600
Range gate max 30000
Range gate min 00000
Timestamp deactivated
Reset LRF module settings command and response
Command
Inputs
Description
>LM,Rd*<CS><CR>
>LM,Rd*7B<0D>
Reset LRF module to default settings.
Example.
Response
Outputs
Description
No response expected.
Table 19: Reset LRF module settings command
Example of a reset LRF module to default settings command transaction:
Host LRF Module
Command frame: >LM,Rd*7B<0D>
LRF Module Host
Acknowledge frame: >AC*84<0D>
Prompt: <