EasyManua.ls Logo

Sanyo MCO-19M - Page 162

Sanyo MCO-19M
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Press the H
2
O
2
Key for 3 seconds. The system check will start.
7. If the system is normal, the following display will appear. (If the system is not normal, refer to Table 2
on page 57.) Select OK and press the ENTER Key to start H
2
O
2
decontamination. Decontamination will
be performed automatically to step 10. The buzzer will sound when decontamination has been completed
(It takes about 1 hour 40 minutes after warming-up).
Note: The outer door will be locked with an electric lock for safety until the decontamination has been
completed.
8. During the H
2
O
2
mist generation, “H2O2 Decon” will flash at the top left corner of the screen.
Note: H
2
O
2
mist generates from the H
2
O
2
Generator and then quickly changes to a gas.
1
H 2 O 2 D ec o n t a m ination
3
4
R e a d y t o S t a rt?
5
6
M E N U
O K
C a n c e l
1
H 2 O 2 D e c o n UV Resolve S t e p:4
2
3
4
5
T e m p : 4 5 . 0
6
D o o r : L o c k e d
7
min
90min
1
H 2 O 2 D e c o n t a m ination
3
4
H 2 O 2 D e c o n S y s tem C h e c k
5
6
- 159 -
52

Table of Contents

Other manuals for Sanyo MCO-19M

Related product manuals