11. UV lamp accumulation time
㻌
㻌
䣫 䣶 䢫䢷 䢰 䣋 䣰䣙 䢻 䢴䢲 䢳 䢰 䣒䢪 䢳䢲䣱 䣰䣈 䣷 䣰 䣥 䣶 䣫
䢲䣅 䣱 䣦 䣧䣏 䣱 䣦 䣧 䣮
䢧䢷 䢵䣨 䣧䣘 䣎 䣫䣗
䢻 䢲䣩 䣑 䢴䣰 䣣 䣮 䣱䣃
䣅 䣣 䣰 䣥 䣧 䣮
䢳 䢳 䢲䢴䣩 䣅 䣑䣰 䣣 䣮 䣱䣃
䣑 䣍
䢲䢳 䢳䣧 䣯 䣲䣱 䣩 䣖䣃 䣰 䣣 䣮
䣏 䣇 䣐 䣗
䣥䣰 䣧 䣶䣣 䣶 䣫 䣱䣮 䣫
䣤 䣴䣩 䣅 䣣䣰 䣣 䣮 䣱䣃
㻌
<Display of UV lamp accumulation time>
UV lamp accumulation time is displayed in the right side of “UV Life” in “Analog Calibration etc” display.
When UV lamp accumulation time reaches 1000hours, it is displayed as 100%.
When a displayed value is “53”%, UV lamp accumulation time is calculated as;
1000hours x 0.53 = 530hours
Note) When UV lamp accumulation time is over 1000hours and UV lamp does not illuminate,
“UV” blinks in the Top Screen.
㻌
㻌
㻌
䣃 䣰 䣣 䣮 䣱 䣩 䣅 䣣 䣮 䣫 䣤 䣴 䣣 䣶 䣫 䣱 䣰 䣧 䣶 䣥
䣏 䣇 䣐 䣗
䣃 䣰 䣣 䣮 䣱 䣩 䣖 䣧 䣯 䣲 䢳 䢲 䢲
䣑 䣍
䣃 䣰 䣣 䣮 䣱 䣩 䣅 䣑 䢴 䢳 䢲 䢲
䣅 䣣 䣰 䣥 䣧 䣮
䣃 䣰 䣣 䣮 䣱 䣩 䣑 䢴 䢳 䢲 䢲
䣗 䣘 䣎 䣫 䣨 䣧 䢲 䢧
䣏 䣱 䣦 䣧 䣮 䣅 䣱 䣦 䣧 䢲
䣈 䣷 䣰 䣥 䣶 䣫 䣱 䣰 䢲 䢪 䢳 䢲 䢳 䢰 䣒
䣙 䢻 䢴 䢷 䢰 䣋 䣰 䣫 䣶 䢫
㻌
㻌
<Reset of UV lamp accumulation time>㻌
When UV lamp is replaced after UV accumulation time reaches 1000hours, it is necessary to reset
UV lamp accumulation time in non-volatile memory.
㻌
Reset of UV lamp accumulation time
Procedures MENU key
0 In Top Screen, press MENU key and select “Tools”. Tools
1 Press ENTER key to display “Select Tools (1/2)” and press MENU key to select “Svc”. Svc
2 Press ENTER key to display “Password” and input service code “384”. OK
3 Press ENTER key to display “Select Tools Svc” and select “Analog Calibration etc”.
4 Press MENU key and select “OK”. OK
5 Press ENTER key to display “Analog Clibration etc” and input “0” in the right side of “UV
Life” column.
6 Press MENU key and select “OK”. OK
7 Press ENTER key to return to “Select Tools Svc” display.