- 56 -
12. Adjustment of heater balance
<Adjustment of balance in bottom heater>
Bottom heater output value can be changed for the prevention of chamber condensation
or when increasing chamber humidity.
Setting range is “0~40”. (Factory default: “10”)
Note) When you input a value over “40”, bottom heater output value is automatically set in “40”.
<Adjustment of balance in door heater>
Door heater output value can be changed for the prevention of condensation on outer door.
Setting range is “0~9”. (Factory default: “6”)
Note) When you input a value over “9”, door heater output value is automatically set in “9”.
