13. Setting of H2O2 decontamination time
<Setting of H2O2 decontamination temperature>
㻌㻌㻌㻌㻌
An apparatus stands by for H2O2 decontamination until it reaches decontamination set temperature +/- 1.0.
Setting range is 37.0~48.0. (Factory default; 45.0)
㻌㻌㻌㻌
㻌㻌㻌㻌Note) When you input a value lower than 37.0, H2O2 decontamination temperature is automatically set in 37.0.
When you input a value higher than 48.0, H2O2 decontamination temperature is automatically set in 48.0.
<Setting of primary H2O2 misting time>
When an apparatus steps to primary H2O2 misting process, ultrasonic oscillator will be active.
Setting range is 1~30 minutes. (Factory default: 7 minutes)
Note) When you input a value lower than “1”, primary H2O2 misting time is automatically set in 1 minute.
When you input a value higher than “30”, primary H2O2 misting time is automatically set in 30 minutes.
Note) When water sensor signal turns off (H2O2 solution s lower than standard) during primary H2O2
misting process, an apparatus steps to secondary resolution process without commenceing
primary H2O2 preservation process and secondary H2O2 misting process.
㻌㻌<Setting of primary H2O2 preserved time>
Setting range is 0~120 minutes. (Factory default: 0 minute)
Note) When you input a value higher than 120, primary H2O2 preserved time is automatically set in 120 minutes.
<Setting of secondary H2O2 misting time>
When water sensor signal turns off (H2O2 solution is lower than standard) or it elapses 15 minutes since secondary
H2O2 misting time expires during secondary H2O2 misting process, an apparatus steps to secondary H2O2
resolution process.
Setting range is 0~30 minutes. (Factory default: 0 minutes)
Note) When you input a value higher than 30, secondary H2O2 misting time is automatically set in 30 minutes.
<Setting of secondary H2O2 resolution time>
Setting range is 30~360 minutes. (Factory default: 90 minutes)
Note) When you input a value lower than 30 minutes, secondary H2O2 resolution time is automatically set in 30minutes.
When you input a value higher than 360minutes, secondary H2O2 resolution time is automatically set in 360minutes.