<Error message in H2O2 decontamination>
(Note Only 1 error message is displayed even if multiple error messages are present.
Error # Status Display Alarm Action
Err 31 When H2O2 generator is
disconnected
“H2O2 Decon System Error”
“Err31: H2O2 Unit No Connect”
Decontamination
mode aborts
Err 32 When H2O2 solution has not
been filled or H2O2 sensor
fails (disconnected)
“H2O2 Decon System Error”
“Err32: H2O2 Level is Low”
Decontamination
mode aborts
Err 33 When a door is open “H2O2 Decon System Error”
“Err32: H2O2 Level is Low”
Decontamination
mode aborts
Err 34 When H2O2 is abnormally
generated
(If H2O2 misting time is over
15minutes than predetermined
time)
(May be caused by faulty
oscillator, faulty fan, poor fan
operation or H2O2 generator
was set incorrectly)
“Err34: H2O2 Volume” Buzzer
sounds
intermittently
Remote
alarm
activates
UV resolution will
be commenced
Err 35 When power fails during H2O2
is generated
“Err35: Power Failed” is displayed
after power returns.
“Decon Failed” is displayed after UV
resolution process is complete.
Buzzer
sounds
intermittently
Remote
alarm
activates
UV resolution will
be commenced
after power
returns
Err 36 When UV lamp fails “Err36: UV Lamp Failer” Buzzer
sounds
intermittently
Remote
alarm
activates
UV resolution
time will be
extended
Err 37 When power fails during UV
resolution process
“Err37: Power Failed” is displayed
after power returns
“Decon Finished” is displayed after
UV resolution process is complete.
UV resolution will
be commenced
again after power
returns