46 Operating Instructions arium
®
mini plus | arium
®
mini
Startup
Equipment damage due to impurities!
If dirt or foreign objects enter the ultrapure water circulation, individual device
components can potentially become clogged or wear out faster.
t If possible, do not touch the cartridge connections. This will prevent impurities
from getting into the device.
t Remove the plugs from the two external cartridge connections.
t Slide the cartridge straight into the guide rails of the cartridge holder on the
front side of the device with the corresponding marking (“L” for ultrapure water
cartridge and “R” for pretreatment cartridge). Make sure to hold the cartridge at
the top when doing this.
t Press the cartridge firmly into the cartridge holder until it locks into the guide
rails with a distinct clicking sound. Check that the cartridge has been installed
securely by turning it slightly.
t Confirm the insertion of the cartridge with the [OK] button.