Operating Instructions arium
®
mini plus | arium
®
mini 69
Cleaning and Maintenance
8.6.3 Changing Cartridges
To replace cartridges, the final filter must be removed and the device depressurized.
Requirements
In the “Change consumable” menu, the menu item [
Pretreatment cartridge], or
[Ultrapure water cartridge] is activated.
y The dialog box “Remove filter” appears.
Procedure
t Remove final filter (see Chapter 8.6.2, page 68).
y The dialog box “Prepare depress.” appears.
t Start depressurization (see Chapter 8.8, page 79)
y When depressurization is complete, the dialog box “Remove cover” appears.
t Remove the front cover.
t Confirm removal of the cover with the [
OK] button.
y The dialog box “Change cartridge” appears.
8.6.3.1 Removing Cartridges
Procedure
t Tap the two protruding terminals of the cartridge together and pull the cartridge
forward and out.
t Confirm removal of the cartridge with the [
OK] button.
y If the menu item [UV lamp] is enabled in the “Change consumable” menu,
the dialog box “Replace the UV lamp” appears.
t If required: Replace the UV lamp
(see Chapter 8.6.5, page 76).
1
Remove cover
5HPRYHWKHIURQWFRYHU
OK
1
Change cartridge
&KDQJHWKHSUHWUHDWPHQWFDUWULGJH5DV
GHVFULEHGLQWKHPDQXDO
OK