Operating Instructions arium
®
mini plus | arium
®
mini 77
Cleaning and Maintenance
8.6.5.1 Removing the UV Lamp
Procedure
t Disconnect the device from the AC power.
y The display goes out. After the power has been restored, the wizard will continue
automatically.
Risk of injury from electrical current and UV radiation!
The UV lamp emits UV radiation and may be live.
t Disconnect the device from the power supply before the old UV lamp is removed.
t Squeeze the metal retaining clip on the old UV lamp and pull it forward to
remove it.
t Remove the metal retaining clip from the cable and store it in a safe place,
e.g. on the magnet securing the cover.
t Disconnect the black connector from the old UV lamp.
t Unscrew the black plastic cover of the UV lamp housing. If required:
Use a suitable tool to help you do this.
t Remove the black plastic cover of the UV lamp housing.
t Carefully pull the old UV lamp out of the lamp housing, package it and safely
dispose of it (see Chapter
10, page 86).