Operating Instructions arium
®
mini plus | arium
®
mini 75
Cleaning and Maintenance
8.6.4.3 Rinsing the Device
Requirements
The new cartridges have been inserted.
Procedure
t If required: Rinse the pretreatment cartridge of the arium
®
mini plus
(see Chapter 6.4.7.3, page 39).
t Prepare rinsing: Connect the dispense tube (see Chapter 6.4.9, page 42)
t Rinsing the Ultrapure Water Cartridge: Start rinsing (see Chapter 6.4.9,
page 42).
y If the bag no longer contains a sufficient amount of water (Tank Fill Level) or
empties during the rinsing process, the bag is filled automatically.
y When rinsing is complete, the dialog box “End rinsing” appears.
y The dialog box “Connect filter” appears..
t Connect final filter (see Chapter 6.4.10, page 42).
t Confirm that the final filter has been connected with the [
OK] button.
y Cartridge replacement is complete.
1
Start rinsing
7DS6WDUWWREHJLQWKHULQVLQJSURFHVV
START
1
Rinsing
PLQ
X
1
Fill bag
0DNHVXUHWKHUHLVVXI¿FLHQWZDWHUWR¿OO
WKHEDJ7DS67$57WR¿OOWKHEDJ
START
1
End rinsing
5HPRYHWKHGLVSHQVHWXEH
OK
1
&RQQHFW¿OWHU
&RQQHFWWKHVWHULOH¿QDO¿OWHUDVGHVFULEHG
LQWKHPDQXDO
OK