Malfunctions
9 Malfunctions
9.1 Error Messages
If an error message is active, dispensing is canceled and locked automatically.
Error message Fault Cause Correction Chapter
Error 0105
Measured values
are not being
displayed.
There is an internal
communications
error.
t Disconnect the device from the AC
power and wait a minute.
t Reconnect the device to the AC power.
7.1, page 56
t If required: Call the Sartorius Service.
Error 0140
Check UV lamp,
replacement may
be needed.
The UV lamp is
not recognized.
The UV lamp is not
connected properly
or defective.
t Check whether the black plug of the UV
lamp is connected correctly.
8.6.5, page 76
t Remove the UV lamp and check for
damage. If required: Change the UV
lamp.
8.6.5, page 76
t If required: Call the Sartorius Service.
Error 0150
Check device
for leakage.
Please contact
your service
technician.
The device has a
leak.
Water has leaked
inside the device.
t Remove the left side cover and check
that the three bag hoses are properly
connected to the device connectors.
If required: Remove the bag and
reconnect it.
8.6.1, page 66
t Drain the tray at the bottom of the
device and check that the connections
are tight during operation.
t If required: Call the Sartorius Service.
Error 0160
Front cover
is open.
Please put the
front cover back
on.
Operation of the
device is not
possible.
The front cover
is not properly
attached.
t Correctly attach the front cover to the
device.
t If required: Call the Sartorius Service.
Error 0163
Check proper
fit of the
pretreatment
cartridge (R).
The pretreatment
cartridge is not
recognized.
The pretreatment
cartridge has not
been properly
inserted.
t Press the pretreatment cartridge in
firmly until you hear a distinct clicking
sound.
t If required:
Remove the cartridge.
Insert cartridge.
t If required: Call the Sartorius Service.
8.6.3.1, page 69
8.6.3.2, page 70
Error 0166
Check proper fit
of the ultrapure
water cartridge
(L).
The ultrapure
water cartridge is
not detected.
The ultrapure water
cartridge has not
been properly
inserted.
t Press the ultrapure water cartridge in
firmly until you hear a distinct clicking
sound.
t If required:
Remove the cartridge.
Insert cartridge.
t If required: Call the Sartorius Service.
8.6.3.1, page 69
8.6.3.2, page 70