Operating Instructions arium
®
mini plus | arium
®
mini 83
Malfunctions
Error 0180
Please contact
your service
technician.
The fill level of
the bag is not
being properly
detected.
The bag is damaged
or improperly
connected.
t Remove the left side cover and check
the bag for damage. If required:
Change bag.
8.6.1, page 66
t Check that the three bag hoses are
properly connected to the device
connectors. If required: Remove the bag
and reconnect it.
8.6.1, page
66
The sensor is
defective.
t Using the tank outlet tubing connected
to “Bag Outlet”, let about 4 of the pure
water out of the bag and check the level
change.
7.2.4, page 58
t If required: Call the Sartorius Service.
Error
Conductivity
measurement of
the ultrapure
water is out of
range.
The conductivity
of the ultrapure
water is outside
of measuring
range.
There is air in the
ultrapure water
circulation.
t Initiate venting. 8.9, page 80
The ultrapure water
cartridge has been
used up.
t Replace the ultrapure water cartridge. 8.6.3, page 63
t To dispense water for testing purposes:
Disable the dispensing lock.
t If required: Call the Sartorius Service.
Error
Temperature
measurement of
the ultrapure
water is out of
range.
The temperature
of the ultrapure
water is outside
of measuring
range.
The temperature of
the ultrapure water
is outside of
measuring range.
t Check if the ambient temperature
complies with the device specifications
in the technical data.
11.2, page 88
t Dispense about 1 liter:
To dispense water for testing
purposes: Disable the dispensing lock.
Dispense and discard about 1 liter of
water.
t If required: Call the Sartorius Service.
Error
Ultrapure water
quality
> XX μS/cm
The limit has
been exceeded
and the
dispensing lock
is preventing
further
dispensing.
There is air in the
ultrapure water
circulation.
t Initiate venting. 8.9, page 80
t Replace the ultrapure water cartridge. 8.6.3, page 63
t To dispense water for testing purposes:
Disable the dispensing lock.
t If required: Call the Sartorius Service.
The limit has
been configured
incorrectly.
t Check the limit. If required: Configure
the limit.