EasyManua.ls Logo

Sartorius arium mini plus - Connecting Feed Water Tubing (Inlet); Connecting Drain Water Tubing (Drain)

Sartorius arium mini plus
97 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
50 Operating Instructions arium
®
mini plus | arium
®
mini
Startup
6.5.7 Connecting Tubing on Device Rear
Contamination of tank filling tubing!
If the tank filling tubing of the arium
®
mini is not stored properly, it may become
contaminated. Sufficient quality of the pure water the bag is filled with is no longer
guaranteed in this case.
t Do not leave the tank filling tubing on the floor or near sources of contamination
(e.g. in the vicinity of an outlet).
t If the tank filling tubing cannot be kept secure while connected to the arium
®
mini: Remove the tank filling tubing and keep it in a safe place.
t If the tank filling tubing is contaminated:
Change the bag (see Chapter 8.6.1, page 66).
Change the tank filling tubing.
t Only use the tank filling tubing to fill the arium
®
mini bag.
Danger of the bag bursting!
If the drain water tubing of the device is sealed off, clogged or exposed to counter-
pressure, the bag may burst.
t Do not seal off, clog or expose the drain water tubing to counter-pressure.
Procedure
t To ensure the tubing securely connected: Use the pre-assembled tubing from
the accessories supplied.
t Connect the tank filling tubing or feed water tubing (see Chapter 6.5.7.1,
page 50).
t Connect the drain water tubing (see Chapter 6.5.7.2, page 51).
t Connect the tank outlet tubing (see Chapter 6.4.7.3, page 39).
6.5.7.1
Connecting the Tank Filling Tubing or Feed Water Tubing (Inlet)
Tip
To prevent the entry of particles into the device, it is recommended to use the
supplied screens.
Procedure
t Connect the tank filling tubing to the connection labeled “Inlet” on the rear of
the device.
t Store the free end of the tank filling tubing in a clean place.
Water leakage!
t If the tubing is deformed or is not inserted deep enough:
After startup, make sure that all external water connections are leak proof!