36
Specifications
Model BA 310 S BA 310 P BA 110
Weighing range structure SuperRange PolyRange
Weighing range/capacity g 310 60/120/310 110
Readability mg 0,001 0,001/0,002/0,005 0,0001
Tare range (by subtraction) g -310 -310 -110
Reproducibility
(standard deviation) g
≤
±0,001
≤
±0,001/0,001/0,003
≤
±0,001
Linearity
g
≤
±0,002
≤
±0,002/0,002/0,005
≤
±0,001
Response time (typical) g
≤
1,5
≤
1,5
≤
2
Adaption to ambient condi-
tions and application require-
ments
←
by selection of 1 of 4 optimized filter levels
→
Display update (depends on
the filter level selected
s
←
0,1 – 0,4 (selectable)
→
Allowable ambient tempera-
ture range
o
C 0 ... +40 0 ... +40 +5 ... +40
Operating temperature range
o
C
←
+10 ... +30
→
Sensitivity drift within
+10 ... +30
o
C
/C
o
≤
±2 ∙ 10
-6
≤
±5 ∙ 10
-6
≤
±2 ∙ 10
-6
Moisture-proof rating
according to DIN 40 040
←
Class F, non-condensing
→
Pan size mm Ø 130 Ø 130 Ø 80
Net weight, approx. kg 4,5 4,5 3,5
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol, bat, MS
Automatic zero tracking
function (can be turned off by
menu code)
←
standard feature
→
Standard calibration weight
(min. accuracy class)
g 200 (F1) 200 (F1) 100 (F1)
Other permissible calibration
weight
(min. accuracy class)
g 100+50, 200+50,
200+100 (F1)
100, 100+50, 200+50,
200+100 (F1)
50 (F1)
AC power source/
power requirements
V~
←
via AC adapter, 230 or 115, -20% ... +15%
→
Frequency Hz
←
48 – 60
→
Power consumption, (typical) VA
←
maximal 16; typical 8
→
Hours of operation with a fully-
charged YRB 02 Z external
battery pack
h
←
48
→
Built-in interface
RS 232 C-S/V24-V28; 7-bit; -even, -odd;
transmission rates 150 … 19200 Baud,
1 or 2 stopbits, software/Hardware handshake