37
Specifications
Model BA 4100 S BA 3100 P BA 2100 S BA 610
Weighing range structure SuperRange PolyRange SuperRange
Weighing range/capacity g 4100 600/1200/3100 2100 610
Readability mg 0,01 0,01/0,02/0,05 0,01 0,01
Tare range (by subtraction) g -4100 -3100 -2100 -610
Reproducibility
(standard deviation) g
≤
±0,01
≤
±0,01/0,01/0,03
≤
±0,01
≤
±0,005
Linearity
g
≤
±0,02
≤
±0,02/0,02/0,05
≤
±0,02
≤
±0,01
Response time (typical) g
≤
1,5
≤
1,5
≤
1,5
≤
1
Adaption to ambient condi-
tions and application require-
ments
←
by selection of 1 of 4 optimized filter levels
→
Display update (depends on
the filter level selected
s
0,1 – 0,4 0,1 – 0,4 0,1 – 0,4 0,1 – 0,6
Allowable ambient tempera-
ture range
o
C
←
0 .. +40
→
Operating temperature range
o
C
←
0 ... +40
→
Sensitivity drift within
+10 ... +30
o
C
/C
o
≤
±2 ∙ 10
-6
≤
±5 ∙ 10
-6
≤
±2 ∙ 10
-6
≤
±5 ∙ 10
-6
Moisture-proof rating
according to DIN 40 040
←
Class F, non-condensing
→
Pan size mm
←
175 x 175
→
Net weight, approx. kg 3,9 3 3 3
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol, bat, MS
Automatic zero tracking
function (can be turned off by
menu code)
←
standard feature
→
Standard calibration weight
(min. accuracy class)
g 2000 (F1) 1000 (F1) 2000 (F1) 500 (F2)
Other permissible calibration
weight
(min. accuracy class)
g 2000+100
2000+2000 (F1)
2000,
2000+1000 (F1)
1000,
1000+500 (F1)
200+100,
200+200,
500+100 (F2)
AC power source/
power requirements
V~
←
via AC adapter, 230 or 115, -20% ... +15%
→
Frequency Hz
←
48 – 60
→
Power consumption, (typical) VA
←
maximal 16; typical 8
→
Hours of operation with a fully-
charged YRB 02 Z external
battery pack
h
←
48
→
Built-in interface
RS 232 C-S/V24-V28; 7-bit; -even, -odd;
transmission rates 150 … 19200 Baud,
1 or 2 stopbits, software/Hardware handshake