38
Specifications
Model BA 210 BA 6100 BA 4100 BA 2100
Weighing range/capacity g 210
6100
4100 2100
Weighing range/capacity mg 0,01 0,1 0,1 0,1
Tare range (by subtraction) g -210 -6100 -4100 -2100
Reproducibility
(standard deviation)
g
≤
±0,005
≤
±0,05
≤
±0,05
≤
±0,05
Linearity
g
≤
±0,01
≤
±0,1
≤
±0,1
≤
±0,1
Response time (typical) g
← ≤
1
→
Adaption to ambient condi-
tions and application require-
ments
←
by selection of 1 of 4 optimized filter levels
→
Display update (depends on
the filter level selected
s 0,1 – 0,6 0,1 – 0,4 0,1 – 0,4 0,1 – 0,4
Allowable ambient tempera-
ture range
o
C
←
0 .. +40
→
Operating temperature range
o
C
←
0 ... +40
→
Sensitivity drift within
+10 ... +30
o
C
/C
o
≤
±10 ∙ 10
-6
≤
±5 ∙ 10
-6
≤
±5 ∙ 10
-6
≤
±10 ∙ 10
-6
Moisture-proof rating
according to DIN 40 040
←
Class F, non-condensing
→
Pan size mm 116 175 x 175 175 x 175 175 x 175
Net weight, approx. kg 1,6 3 3 1,6
Selectable weight units g, kg, ct, lb, oz, ozt, tlh, tls, tlt, GN, dwt, mg, /lb, tlc, mom, K, tol, bat, MS
Automatic zero tracking
function (can be turned off by
menu code)
←
standard feature
→
Standard calibration weight
(min. accuracy class)
g 100 (M1) 5000 (F2) 2000 (F2) 1000 (M1)
Other permissible calibration
weight
(min. accuracy class)
g 100+50,
200 (M1)
2000+1000,
2000+2000,
2000 (F2)
2000+1000,
2000+2000 (F2)
1000+500;
2000 (M1)
AC power source/
power requirements
V~
←
via AC adapter, 230 or 115, -20% ... +15%
→
Frequency Hz
←
48 – 60
→
Power consumption, (typical) VA
←
maximal 16; typical 8
→
Hours of operation with a fully-
charged YRB 02 Z external
battery pack
h
←
48
→
Built-in interface
RS 232 C-S/V24-V28; 7-bit; -even, -odd;
transmission rates 150 … 19200 Baud,
1 or 2 stopbits, software/Hardware handshake