Operating the Scale
Line Format (Data Output Format)
You can output the values
displayed in the measured value line
and the weight unit with or without
a data ID code
Example: Without data ID code
+ 253 pcs
Example: With data ID code
Qnt + 253 pcs
Configure this parameter in the Setup
menu under item 7 2.
The output with data ID code has
16 characters; without data ID code,
22 characters.
Output Format with 16 Characters
Display segments that are not activated are output as spaces.
Characters without a decimal point are output without a decimal point.
The following characters can be output, depending on the characters
displayed on the scale:
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+*DDDDDDDD*UUUCRLF
or ........ ***
or * ********
or 000000
*: Space
D: Digit or letter
U: Unit symbol
CR: Carriage return
LF: Line feed
Special Codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
************CRLF
or H H
or L L
or C
*: Space
– –: Weight; all numbers shown in stable readout
H: Overload
H H: Overload during checkweighing
L: Underload
L L: Underload during checkweighing
C: Calibration/adjustment
Error codes
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
***E r r *###****CRLF
*: Space
# # #: Error code number