67
Operating the Scale
Data Input Format
You can connect a computer to your scale to send commands via the scale
interface port to control scale functions and applications.
The commands sent are control commands and may have different formats;
e.g., control commands can have up to 26 characters. Each character must
be transmitted according to the settings configured in the Setup menu for
data transmission.
Farmat for Control Commands
Format 1: Esc ! CR LF
Format 2: Esc ! # _ CR LF
Format 3: Esc ! # & (max. 20 &) & _ CR LF
Format 4: Esc ! _ CR LF
Esc: Escape _: Underline (ASCII: 95)
!: Command character CR: Carriage return
#: Number LF: Line feed
&: Number or letter max: Depends on command
character; i.e., parameter:
See under “Overview” once the max. length
for an ASCII table is reached, input received
via the interface port is
cut off, rather than rejected
as for a key input
Format 1
! Meaning
K Weighing mode 1
L Weighing mode 2
M Weighing mode 3
N Weighing mode 4
O Block keys
P Print
R Unblock keys
S Restart
T Tare t
Z Internal calibration/adjustment
Format 2
!# Meaning
f0 Function key w/[TOGGLE]
f1 Function key [ISOTEST]/[TEST]
f2 Function key v/[FUNCTION]
f3 Zero (separate zeroing key)
f4 Tare (separate zeroing key)
s3 Function key c
x0 Perform internal calibration
x1 Print scale model
x2 Print weighing cell serial number
x3 Print weighing cell
software version
x4 Print display and control unit
software version
x5 Print scale ID number
x6 Print weight set number
x7 Print lot number
(weighing series ID)
Format 3
(not allowed in the Setup menu)
!# Meaning
z5 Input scale ID number
z6 Input weight set number
z7 Input lot number
Format 4
! Meaning
t Text input in display