Step  Key (or  instruction)    Display/output
9.  Open the hinged cover  Open the LMA200PM    Prompt (green background):
  Remove the pads    Add sample to pad
10. Add sample  Trickle ketchup onto glass fiber pad
    Spread ketchup evenly
    Place the second pad on the ketchup
11. Insert sample  Place pads with ketchup on the     If the sample weight is within the 
    sample retainer    tolerance limits:
        An acoustic signal (beep) is emitted
        The bar graph is green
        Close cover or press Start
        If the sample weight is not within 
        tolerance limits:
        No acoustic signal is emitted
        The bar graph is orange or red
  Change sample amount as needed  Remove some ketchup from the pad, or
    add more ketchup
12. Continue analysis  Close the analyzer cover    The status line briefly shows 
        the following:
        Determining initial weight value...
        The header is printed. Example:
      –––––––––––––––––––––––––––––––––––––––––
      
28.11.2007  16:41:46
   
Sartorius Microwave Moisture Analyzer
       LMA200PM-000EU  Rel. 01.00.00
      Serial no.  12345678
      –––––––––––––––––––––––––––––––––––––––––
      Prg.        Ketchup
      Mode: Standard | Process: R Standard
      Level:    30 %
      End:  Automatic
      Initial weight: +    2.125 g
      –––––––––––––––––––––––––––––––––––––––––
        During analysis, the status line shows 
        the following (yellow background):
        Analysis in progress...
         At the end of analysis, the status line briefly 
shows the following (yellow background):
        Determining final weight...
        The analysis result is printed. Example:
      –––––––––––––––––––––––––––––––––––––––––
      00:40 + 0.752 g + 64.63 %M
      –––––––––––––––––––––––––––––––––––––––––
      After the end of analysis, the prompt line
        shows the following (green background):
        Run completed; remove sample
      The bar graph line shows the percentage 
        of solids in the initial weight. 
        The measured value line shows, for example:
        Moisture    + 64.63 %M
13. Remove the sample  Raise the analyzer cover    The cooling fan continues to run.
  Remove the pads with the sample.  The cooling fan shuts off automatically.
14. Begin next analysis  Tap the Analyze button    Continue from Step 4 above.
        Cooling fans switch off.
15. End the analysis series.   Press the 
 
key    Starting screen
26