105
Specifications
User Manual Secura
®
, Quintix
®
, Practum
®
15.2 Model-specific Data
Secura
®
Model
224-1x
1)
124-1x
1)
1103-1x
1)
613-1x
1)
513-1x
1)
313-1x
1)
213-1x
1)
6102-1x
1)
5102-1x
1)
3102-1x
1)
2102-1x
1)
1102-1x
1)
612-1x
1)
6101-1x
1)
3101-1x
1)
Weighing capacity g 220 120 1,100 610 510 310 210 6100 5100 3100 2100 1100 610 6,100 3,100
Readability mg 0.1 0.1 1 1 1 1 1 10 10 10 10 10 10 100 100
Repeatability
(standard deviation)
mg 0.1 0.1 1 1 1 1 1 10 10 10 10 10 10 50 50
Linearity deviation mg 0.2 0.2 2 2 2 2 2 20 20 20 20 20 20 100 100
Min. initial weight according
to USP, typical
g 0.12 0.12 1.5 1.5 1.5 1.5 1.5 12 12 12 12 12 12 60 60
Sensitivity drift
between +10 to +30°C
± ppm/K 1.5 1.5 1.5 2 2 2 2 2 2 2 2 2 2 2 2
Typical stabilization time s 2 2 1.5 1 1 1 1 1 1 1 1 1 1 1 1
isoCAL:
Temperature change
Time interval
K
h
1.5
4
1.5
4
1.5
4
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
Display result (depending
on the set filter level)
s 0.2 0.2 01/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
0.1/
0.2
Weighing pan size mm
d 90 d 90 d 120 d 120 d 120 d 120 d 120 d 180 d 180 d 180 d 180 d 180 d 180 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209 209
Net weight, approx. kg 5.1 5.1 5.9 5.1 5.1 5.1 5.1 5.2 5.2 5.2 5.2 5.2 4.7 5.2 5.2
Quintix
®
Model
224-1x
1)
124-1x
1)
64-1x
1)
613-1x
1)
513-1x
1)
313-1x
1)
213-1x
1)
6102-1x
1)
5102-1x
1)
3102-1x
1)
2102-1x
1)
1102-1x
1)
Weighing capacity g 220 120 60 610 510 310 210 6100 5100 3100 2100 1100
Readability mg 0.1 0.1 0.1 1 1 1 1 10 10 10 10 10
Repeatability
(standard deviation)
mg 0.1 0.1 0.1 1 1 1 1 10 10 10 10 10
Linearity deviation mg 0.1 0.1 0.1 2 2 2 2 20 20 20 30 30
Sensitivity drift
between +10 to +30°C
± ppm/K 1.5 1.5 1.5 3 3 3 3 3 3 3 4 4
Typical stabilization time s 2 2 2 1 1 1 1 1 1 1 1.5 1.5
isoCAL:
Temperature change
Time interval
K
h
1.5
4
1.5
4
1.5
4
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
2
6
Display result (depending
on the set filter level)
s 0.2 0.2 0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2 0.1/0.2
Weighing pan size mm
d 90 d 90 d 90 d 120 d 120 d 120 d 120 d 180 d 180 d 180 d 180 d 180
Weighing chamber height* mm 209 209 209 209 209 209 209
Net weight, approx. kg 4.9 4.9 4.9 4.9 4.9 4.9 4.9 5.2 5.2 5.2 4.7 4.7
1
) Possible terms for country-specific models:
x = S: Standard balances without country-specific additions
x = SAR: Standard balances with country-specific additions for Argentina
x = SJP: Standard balances with country-specific additions for Japan
x = SKR: Standard balances with country-specific additions for South Korea
* Upper edge of the weighing pan to the lower edge of the upper draft shield panel