USB Port
User Manual Secura
®
, Quintix
®
, Practum
®
Special Outputs
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * * * * * * * * * * CR LF
or * * * * * * L o w * * * * * CR LF
or * * * * * * H i g h * * * * CR LF
* Space Cal.Ext. External adjustment
– – Final readout Cal.Int. Internal adjustment
Low Underweight CR Carriage return
High Overload LF Line feed
Error message
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * E r r * */# # # * * * * CR LF
* Space ### Error number
Example: Output of the weight value + 111.255 g
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
* * * 1 1 1 . 2 5 5 * g * * CR LF
+ * * 1 2 3 . 5 [ 6 ]
1)
g * * CR LF
Position 1: Plus +, minus –, or space
Position 2 – 10: Space or weight value with decimal point;
leading zeros are output as spaces.
Position 11: Space
Position 12 – 14: Characters for unit of measure or space
Position 15: Carriage return
Position 16: Line feed
1)
Settings “PC – SBI” and “USB RS232 SBI” for the identification of digits not verified
for use in legal metrology:
In the setting “SBI”, there is no automatic identification of display digits not
verified for use in legal metrology. Corresponding measures or settings must be
carried out on the peripheral device.