Technical Specifications
Model-specific Data
Verified Models with Country-specific Type Approval Certificate
Secura
®
Model 26-1x 
1)
225D-1x 
1)
125-1x 
1)
324-1x 
1)
Accuracy class
k k k k
Type 
2
) SQP-H SQP-F SQP-F SQP-G
Max g 21 120 / 220 120 320
Scale interval d g 0.000002 0.00001 / 0.0001 0.00001 0.0001
Verification scale interval e g 0.001 0.001 0.001 0.001
Min g 0.001 0.001 0.001 0.01
Tare (subtractive) < 100% of max. weighing capacity
Minimum sample weight  
as per USP, typical
mg 4 40 40 160
Typical stabilization time s 8 6 / 2 6 2
isoCAL: 
– Temperature change 
– Time interval
 
K 
h
 
1.5 
4
 
1.5 
4
 
1.5 
4
 
1.5 
4
Display result  
(depending on the set filter 
level)
s 0.2 / 0.4 0.2 / 0.4 0.2 / 0.4 0.2 / 0.2
Weighing pan size mm
d 50 d 80 (optional d 90) d 80 (optional d 90) d 90
Weighing chamber height* mm 218 218 218 218
Net weight, approx. kg 8.0 7.8 7.8 7.9
Quintix
®
Model 125D-1x 
1)
65-1x 
1)
35-1x 
1)
Accuracy class
k  k k
Type 
2
) SQP-F SQP-F SQP-F
Max g 60 / 120 60 30
Scale interval d g 0.00001 / 0.0001 0.00001 0.00001
Verification scale interval e g 0.001 0.001 0.001
Min g 0.001 0.001 0.001
Tare (subtractive) < 100% of max. weighing capacity
Typical stabilization time s 6 / 2  6 6
isoCAL: 
– Temperature change 
– Time interval
 
K 
h
 
1.5 
4
 
1.5 
4
 
1.5 
4
Display result  
(depending on the set filter 
level)
s 0.2 / 0.4 0.2 / 0.4 0.2 / 0.4
Weighing pan size mm
d 80 (optional d 90) d 80 (optional d 90) d 80 (optional d 90)
Weighing chamber height* mm 218 218 218
Net weight, approx. kg 7.8 7.8 7.8
1)
  Possible terms for country-specific models:
  x = CEU: Verified balances with EC Type Approval Certificate  
  D12-09-014 without country specific additions
  x = CFR: Verified balances with EC Type Approval Certificate  
  D12-09-014 for France only
  x = CIT: Verified balances with EC Type Approval Certificate  
  D12-09-014 for Italy only
  x = CCH: Verified balances with EC Type Approval Certificate  
  D12-09-014 for Switzerland only
  x = NUS: NTEP Certificate for USA
  x = CN: CMC Type Approval Certificate for China
  x = 0JP: Balance with Type Approval Certificate for Japan
  x = 0BR: Balance with Type Approval Certificate for Brazil
  x = 0RU: Balance with Type Approval Certificate for Russia
  x = 0IN: Balance with Type Approval Certificate for India
  x = 0CA: Balance with Type Approval Certificate for Canada
2
)   All models with “...CN”: type “SQP”
*   Upper edge of the weighing pan to the lower edge of the upper 
draft shield panel
  Installation and Operating Instructions Manual Secura
®
, Quintix
®