8. Technical Data
General Data
(Model-specific data see page 80)
Information Unit Value
AC Adaptor
Sartorius AC adaptor module
Primary 100 –240 V~, –10% / +10%, 50–60 Hz, 0.2 A
Secondary 15 V DC, ± 5%, 530 mA (max.) / 8 W (max.): 0 to +40 °C
15 V DC, ± 5%, 330 mA (max.) / 5 W (max.): 0 to +50 °C
Further data Protection class II in accordance with EN / IEC 60950-1 | up to 3000 m above
sea level
IP40 in accordance with EN 60529 / IEC 60529
Balance
Power supply Only via Sartorius AC adaptor module 6971790
Input voltage V DC 12.0 – 15.0
Power consumption W 4.5 (typically)
Further data IP43 in accordance with EN 60529 / IEC 60529
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment For indoor use only
Ambient temperature* °C +10 – +30
Operational capability °C Guaranteed between +5 – +45
Storage and shipping °C -10 – +60
Elevation m above
sea level
Up to 3000
Relative humidity** % 15 – 80% for temperatures up to 31 °C non-condensing, decreasing linearly
up to 50% relative humidity at 40 °C and 20% at 50 °C.
Safety of Electrical Equipment As per EN 61010-1 / IEC61010-1 Safety requirements for electrical equipment
for measurement, control and laboratory use – Part 1: General requirements
Electromagnetic Compatibility As per EN 61326-1 / IEC61326-1 Electrical equipment for measurement,
control and laboratory use – EMC requirements – Part 1: General requirements
Defined immunity to interference Suitable for use in industrial areas
Interference emission Class B
Suitable for use in residential areas and areas that are connected to a low
voltage network that also supplies residential buildings.
Standard Equipment
Selectable weight units
1
Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong teal, Singapore
teal, Taiwan teal, grain, pennyweight, milligram, parts per pound, China teal,
mommas, Austrian carat, tola, baht, mesghal and Newton
Available application programs Secura
®
Models
Counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing, mixing, components (totalizing), statistics,
conversion, SQmin and identifiers
Quintix
®
Models
Counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing, mixing, components (totalizing), statistics,
conversion
Verified balances in accordance with EU requirements comply with the requirements of the Council Directive 2009/23/EG with
EN 45501:1992 and OIML R76:2006.
* For verified balances in accordance with EU requirements, refer to the information on the balance.w
** For verified balances in accordance with EU requirements, the legal regulations apply.
1
Depending on the country-specific model version, not all weight units listed may be available.
Installation Instructions Secura, Quintix 9