EasyManua.ls Logo

Sartorius Secura Quintix 26 - Datos Específicos del Modelo; Dati Specifici del Modello; Сведения О Модели; 具体型号数据

Sartorius Secura Quintix 26
83 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Specifications
Model-specific Data
Secura
®
Model 26-1x
1)
225D-1x
1)
125-1x
1)
324-1x
1)
Weighing capacity g 21 60 / 120 / 220 60 / 120 320
Readability mg 0.002 0.01 / 0.01 / 0.1 0.01 / 0.01 0.1
Repeatability (standard deviation) mg 0.004 0.03 / 0.04 / 0.07 0.03 / 0.04 0.1
Repeatability (standard deviation), typical mg 0.003 0.02 / 0.04 / 0.07 0.02 / 0.04 0.1
Linearity mg 0.01 0.1 / 0.1 / 0.2 0.1 / 0.1 0.3
Min. initial weight
according to USP, typical
mg 4 40 40 160
Sensitivity drift between +10 to +30°C ± ppm/K 1 1 1 1
Typical stabilization time s 8 6 / 6 / 2 6 2
isoCAL Settings:
– temperature change
– time interval
K
h
1.5
4
1.5
4
1.5
4
1.5
4
Display result (depending
on the filter level)
s 0.2 / 0.4 0.2 / 0.4 0.2 / 0.4 0.2 / 0.2
Weighing pan size mm
d 50 d 80 (d 90
optional)
d 80 (d 90
optional)
d 90
Weighing chamber height* mm 218 218 218 218
Net weight, approx. kg 8.0 7.8 7.8 7.9
Quintix
®
Model 125D-1x
1)
65-1x
1)
35-1x
1)
Weighing capacity g 40 / 60 / 120 40 / 60 30
Readability mg 0.01 / 0.01 / 0.1 0.01 / 0.01 0.01
Repeatability (standard deviation) mg 0.03 / 0.04 / 0.07 0.03 / 0.04 0.03
Repeatability (standard deviation), typical mg 0.02 / 0.04 / 0.07 0.02 / 0.04 0.3
Linearity mg 0.1 / 0.1 / 0.2 0.1 / 0.1 0.1
Sensitivity drift between +10 to +30°C ± ppm/K 1 1 1
Typical stabilization time s 6 / 6 / 2 6 6
isoCAL Settings:
– temperature change
– time interval
K
h
1.5
4
1.5
4
1.5
4
Display result (depending
on the filter level)
s 0.2 / 0.4 0.2 / 0.4 0.2 / 0.4
Weighing pan size mm
d 80 (d 90 optional) d 80 (d 90 optional) d 80 (d 90 optional)
Weighing chamber height* mm 218 218 218
Net weight, approx. kg 7.8 7.8 7.8
FCC Rules
Safety Precautions:
NOTE: This equipment has been tested and
found to comply with the limits for a Class B
digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide
reasonable protection against harmful
interference in a residential installation. This
equipment generates, uses and can radiate
radio frequency energy and, if not installed
and used in accordance with the instructions,
may cause harmful interference to radio com-
munications. However, there is no guarantee
that interference will not occur in a particular
installation. If this equipment does cause
harmful interference to radio or television
reception, which can be determined by
turning the equipment off and on, the user is
encouraged to try to correct the interference
by one or more of the following measures:
Reorient or relocate the receiving
antenna.
Increase the separation between the
equipment and receiver.
Connect the equipment into an outlet on
a circuit different from that to which the
receiver
Consult the dealer or an experienced
radio/TV technician for help.
Shielded Cables:
Connections between the devices and periph-
erals must be made using shielded cables in
order to maintain compliance with FCC radio
frequency emission limits.
Modifications:
Any modifications made to this devices that
are not approved by Sartorius Weighing
Technology GmbH may void the authority
granted to the user by the FCC to operate this
equipment.
1)
Possible terms for country-specific models:
x = S: Standard balances without country
specific additions
x = SAR: Standard balances with country
specific additions for Argentina only
x = SJP: Standard balances with country
specific additions for Japan only
x = SKR: Standard balances with country
specific additions for South Korea only
* Upper edge of the weighing pan to the
lower edge of the upper draft shield panel
80 Installation and Operating Instructions Manual Secura
®
, Quintix
®

Table of Contents