26
Aut12420.jpg
Activating the BPI Mode
Note:
The BPI mode must be activated in the balance/scale before you can work with
the service software. You need to use this software, for example, to adjust the
linearity and/or span after adjusting the zero-point offset value, or after
replacing a PCB.
The BPI adapter (part no. 6740-71) must be connected to the balance/scale
before the BPI mode can be activated. The procedure for connecting the
adapter is as follows:
Turn off the balance/scale and disconnect it from the power supply
Remove the weighing pan/load plate
Remove the protective cap (A) from the bottom of the balance/scale.
The main PCB is now accessible through an opening in the balance/scale
housing.
Connect the white plug on the cable of BPI adapter to the corresponding
position on the main PCB.
Connect the plug on the power cable of the BPI adapter to the power jack
on the balance/scale.
Plug the balance/scale AC adapter into the power jack on the BPI adapter.
BPI-Button
Anschluss
BPI02.jpg
BPI-Button
Main PCB