EasyManua.ls Logo

Sartorius Talent TE153S - Linearity Testing and Adjustment (Double Lever System); Checking and Adjusting Linearity

Sartorius Talent TE153S
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
47
Adjusting Linearity
Checking the Linearity
Note: The linearity must be checked according to the Standard Operating Procedure
WKD-038.
- Check the linearity of the balance/scale for the entire weighing range in 4-g
steps.
- Compare the values displayed with the tolerance ranges given in the „Table
of Adjustment Data“ Page 53-54.
- If the errors exceed the permissible tolerance ranges, the linearity must be
adjusted.
Adjusting the Linearity
Use the service software to adjust linearity.
Activate the BPI mode (see page 26).
Connect the balance/scale to the PC and call the linearity program in the
service software.
Test the span and adjust if necessary (see page 46, „Span Adjustment“).
After linearity has been adjusted, activate the „Close“ function in the service
software to reset the balance/scale to the SBI mode.

Related product manuals