EasyManua.ls Logo

SBM IDRA - RF Wireless Communication Immunity; Immunity Test: RF Wireless Communication

SBM IDRA
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
25
Port
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Frequency
(MHz)
385
450
710
745
780
810
870
930
1720
1845
1970
2450
Polariz.
antenna
V/O
V/O
V/O
V/O
V/O
V/O
V/O
V/O
V/O
V/O
V/O
V/O
Level (V/m)
27
28
9
28
28
28
Reference
Standard
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
Immunity to rf wireless communication equipment
Reference document:
Test level:
Test Equipment:
Environmental Conditions:
Uncertainty:
EN 61000-4-3
see table below
Generator: Rhode & Schwarz SM300; Amplifier: Kalmus 737FC;
Frankonia Amplifier FLG30C; Log- periodic antenna Frankonia
BTA-M; Management software Frankonia; Isotropic sensor
PMM EP 330
Temperature 20°C, RH 40%
±3.26 (dB)
Pulse
Modulation
18 Hz
18 Hz
217 Hz
18 Hz
217 Hz
18 Hz