EasyManua.ls Logo

SBM IDRA - Magnetic Field Immunity and Test Summary; Immunity Test: Power Frequency Magnetic Fields; Summary of All Test Results

SBM IDRA
32 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26
Port
Enclosure
Range freq. (Hz)
50
Level (A/m)
30
Reference Standard
EN 60601-1-2
Power frequency magnetic fields immunity test
Reference document:
Level:
Test Equipment:
Environmental Conditions:
Uncertainty:
EN 61000-4-8
EN 60601-1-2
Generator PMM DAS-G120, Coil DANA DAS S120
Temperature: 20°C, RH: 40%
<8%
SUMMARY OF TESTS RESULTS
The following table summarizes the tests performed:
Port
Enclosure
Enclosure
Enclosure
Enclosure
Enclosure
Frequency range
(MHz)
30-1000
.............
80-2700
See point 6.2.3
0.00005
Test Level
Class B
±15KV air
±8KV contact
3V/m modulated
AM 80% 1KHz
See point 6.2.3
30A/m
Test Type
Radiated emission
Electrostatic discharge
immunity
Radiated electromagnetic
field immunity test
Immunity to RF wireless
communication
equipment
Power frequency
magnetic fields
immunity test
Reference
Standards
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
EN 60601-1-2
Result
Ok
Ok
Ok
Ok
Ok