EasyManua.ls Logo

SCHOTT Instruments ProLab 2000 - Page 77

SCHOTT Instruments ProLab 2000
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProLab 2000 Operation
77
ba75577e03 11/2009
1 Put the D.O. sensor into the air calibration vessel.
2 Connect the D.O. sensor to the meter.
3 In the measured value display, select the D.O. measuring
window with <><> and <MODE>.
4 Start the calibration with <CAL>.
The calibration display for the relevant sensor type appears.
5 Put the D.O. sensor into the air calibration vessel.
6 Press <MENU/OK>.
The measurement with stability control to determine the
relative slope starts.
7 Wait for the end of the measurement with Stability control or
accept the calibration value with <MENU/OK>.
The determined relative slope is displayed as a message (
i).
i Relative slope = 1.00
i Use air calibration vessel
Continue
LFOx1400-ID <CAL>
i Relative slope = 1.00
i Temperature =24.8 °C
Terminate stability control
LFOx1400-ID <CAL>

Table of Contents