EasyManua.ls Logo

SCHOTT Instruments ProLab 2000 - Page 78

SCHOTT Instruments ProLab 2000
125 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation ProLab 2000
78
ba75577e03 11/2009
Calibration via a
Comparison meas.
For this calibration procedure, the Comparison meas. setting must be
set to On in the measuring menu.
Note
Before calibrating via a comparison measurement, the sensor should
be calibrated in the air calibration vessel.
Proceed as follows to calibrate the meter:
8 You have the following options:
z Accept the new calibration values with <MENU/OK>.
Subsequently, the calibration record is displayed and
downloaded to the interface at the same time.
z To switch to the measured value display without
accepting
the new calibration values, press <MODE> or <ESC>.
i Relative slope = 0.98
i Sensor+++
Accept
LFOx1400-ID <CAL>
1 Connect the D.O. sensor to the meter.
2 Immerse the D.O. sensor in the reference solution.
3 In the measured value display, select the D.O. measuring
window with <><> and <MODE>.
4 Start the calibration with <CAL>.
The calibration display appears.
i Immerse sensor in ref. sol.
Continue
LFOx1400-ID <CAL>

Table of Contents