ProLab 2000 Operation
79
ba75577e03 11/2009
5 Press <MENU/OK>.
The measurement with stability control to determine the D.O.
concentration starts.
6 Wait for the end of the measurement with Stability control or
accept the measured value with <MENU/OK>.
The determined D.O. content is displayed as a message (
i).
7 Open the setting of the correction factor with <MENU/OK>.
8 Using <><>, set the correction factor to adjust the
displayed concentration value to the nominal value (value of
the comparison measurement). Subsequently, accept the
correction factor with <MENU/OK>.
9 Select Accept with <><> and press <MENU/OK>.
The meter switches to the measured value display. The
[Factor] indicator appears in the measuring window.
i Concentration = 8.5 mg/l
i Temperature =24.8 °C
Terminate stability control
LFOx1400-ID <CAL>
i Concentration = 8.49 mg/l
i Temperature =24.8 °C
Set factor: 1.000
Accept
LFOx1400-ID <CAL>