the load cell size entered during start-up. Make sure that you use the same no. of digits for both
“Val Now” and “New Val” before initiating “Change”.
Example 2, adjustment with material test:
- Check that the display flowrate is indicating 0. If not, initiate an auto zero sequence or a system
tare out with “SysZERO”.
- Zeroset a desired totaliser to use for the test (Sub1Tot, Sub2Tot).
- Let a known/verified quantity of material pass over the belt scale.
- Enter the indicated totalised quantity of test material as value for parameter “Val Now”.
- Enter the verified quantity of the same material as value for parameter “New Val”.
- Initiate change by selecting “Yes” for parameter “Change”.
Adjustment has now been calculated and changed for which can be seen as a chnage of parameter
“Cur.Adj”in%.
Example 3, adjustment with an already known/calculated devation in %:
- Enter value “1000” for parameter “Val Now” as current value.
- Enter value “1015” for parameter “New Val” for a desired +1,5% adjustment.
- Initiate change by selecting “Yes” for parameter “Change”.
You will now find that parameter “Cur.Adj” has changed to 101,5% as current value, resulting in a
flowrate and totalising increase of 1,5%.
TIME SETUP ->
This menu contains parameters for date and time settings. The internal clock will continue working when
power is disconnected, as long as the internal back-up battery is working.
Year= 2006 - 2199
Month= 1-12
Day= 1-31
Hour= 0-23
Minute= 1-59
TimeSet Yes/No. Resets the clock with entered values above.
COM SETUP ->
This menu contains control parameters for the RS485 serial port.
BPS= Baudrate, bits/s
Selectable between: 1200 - 38400
Default setting: 9600
Format= Selectable between:
“Data xyz” for transmit/receive communication to PC, PLC etc.
“Prn xyz” for transmitting only (no handshake or contol signals) to printer
or external remote display unit.
x = No. of bits: 7 or 8
y = Parity: none, even or odd
z = Stop bit: 1 or 2
Default setting: Data 8n1