The pump doses a programmed quantity in mL. It is possible to set a
delay time (Delay) for the pump start up and an interval between
subsequent dosages (Interval) as shown on the diagram:
The Delay and Interv. times are in dd.hh.mm format
(days.hours.minutes)
The Pause input can be programmed in three different modes:
1. Freeze Time: with the pause activated, the system stops the
current time count and restarts the count when the pause is
deactivated
2. Pause Dosing: with the pause activated, the system
continues to count the time and stops the dosage.
3. Restart Timer: with the pause activated, the system stops the
dosage and when the pause is deactivated the count restarts
from the beginning.
The dosage frequency can be modified during the pump operation by
pressing
+ simultaneously.