172
7. Meter Settings [Setting] Screen
When the Following Screen is Displayed
The Light Selection Ring is set to Range H
( ) or Range L ( ), and dark calibration
could not be performed.
Set the Light Selection Ring
to the dark calibration
position
( ), and perform dark calibration.
While the dark calibration is being performed, the
position of Light Selection Ring
is changed. so
dark calibration could not be completed normally.
Set the Light Selection Ring
to the dark calibration
position
( ) again, and perform dark
calibration again.
Dark calibration can be done by turning the
Light Selection Ring
to set to the dark
calibration position
( ) in Measuring
Screen and Display Mode Selection screen.
Dark calibration is performed when there is a
big change in temperature between turning
power OFF and ON.
Except the cases above, dark calibration
after power ON is skipped.
Light Selection Ring
Dark Calibration Position
Dark Calibration
failure Screen