6 . ERROR MESSAGES
2000
OpEN
Err1
----
OL
Measure in overrange.
Sensor broken.
Upper or lower electrical overstepping of the input.
Displayable value overload.
7 . GENERAL WARRANTY TERMS
WARRANTY APPLYING AND DURATION:
This appliance is guaranteed for 1 year for any design or manufacturing defects, under normal operating conditions.
Conditions for processing *:
Processing not under warranty will be submitted to the acceptance of a repair estimate.
The customer will return the products at his charge, and they will be restored to him after processing.
Without a written agreement on the repair estimate within 30 days, the products will not be kept.
* Complete warranty terms and details available on request.
Er
Value set out of range.
Self-diagnosis error.
Coding:
Err. 1: Programming error (programming parameters incoherent).
Err. 4: Error on internal offset (excessive drift).
Err. 8: Calibration error.
Err. 32: Error on the CJC (excessive drift).
Err. 64: Upper or lower electrical overstepping of the input.
If the instrument detects for example an offset error (4) and a programming error (1) the
value of the error code will be 5 (4+1).