Chapter 4 Method Parameters
Start delay time/Extraction time (KF):
With dead-stop titration, the start wait time passes at the beginning of titration. In KF titration, the start wait time =
the extraction time. The extraction time ends after the sample is supplied. The start wait/extraction time can be
specified between 0 and 999 seconds.
Fig. 88
Conditioning
Conditioning (only KF) is activated for every KF method. It can be shut off via a PC for external control:
Fig. 89
Fixed delay time
The fixed delay time is the waiting time between the linear titration steps at the end of the titration until the
Endpoint. The fixed delay time can be set between 0 and 999 seconds:
Fig. 90
Step size
The step size can be set from 0.001 to 5.000 ml. typical value for the KF titration are 0.002 0.01 ml.