Chapter 4 Method Parameters
Fig. 94
The values can be set between 40 and 220 m. The pre-setting is 100 mV.
Low polarisation voltage: insensitive
High polarisation voltage: sensitive
4.6.4 Titration end
The end of a titration is reached, and the result will be calculated as soon as, or if, respectively:
The defined End value in µA value has been reached
The Endpoint delay in seconds has been adhered
The drift value in µg/min has been reached
The predefined value ml has been reached (Maximum titration volume)
The conditions for minimum and maximum titration time are maintained
Fig. 95
Fig. 96
Maximum titration time
The maximum titration time can be set between 0 9999 seconds. The default setting is 600 seconds. The
maximum titration time is generally used for KF titration, which can create a high continuous drift from a
secondary reaction and thus cannot reach a stable endpoint.