Technical Information
2.2 Design
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
15
Inputs and outputs
One to three measurement channels with hybrid connections for the sensors at the
measuring points
Two analog inputs per channel for process gas temperature and pressure
Two analog outputs per channel for gas concentration(s) or for gas temperature and
concentration For selected versions, the transmission can be read out as an alternative.
Six freely configurable binary inputs per channel for signalling faults or maintenance
requests from external temperature or pressure transducers or sensor purging failure.
Six freely configurable binary outputs per channel (signalling of faults, maintenance
requirements, function control, transmission limit alarms, concentration limit alarms, store
analog output)
Communication
Network connection: Ethernet (10Base-T) for remote diagnostics and maintenance.
The LDS 6 can be operated remotely via the Ethernet port with a PC running
Windows 95/98/ME or Windows NT/2000/XP. It is also possible to connect the LDS 6 via
modem to the public telephone net. In that case an LDS 6 LAN modem kit is required. Any
external connection requires the optional software LDSComm (LDS Communication Client)
to be installed on the remote computer. All aspects of LDS 6 can be controlled in this way.
For the operation of the LDS 6 using LDSComm software refer to the LDSComm Manual
(A5E02183317).
Maintenance and fault messages
LDS 6 outputs different warnings via relays:
Need for maintenance (measured value is not influenced)
Operating error (measured value might be influenced)