Siemens LDS 6 - 8.2 Hybrid and Sensor Cables

Siemens LDS 6
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Data
8.2 Hybrid and Sensor Cables
LDS 6
Operating Instructions, 01/2009, A5E00295894-05
87
8.2 Hybrid and Sensor Cables
Table 8- 10 General
General
Configuration hybrid cable Two optical fibers and two twisted copper wires in
one cable for 24 V DC. Single-mode optical fiber
configured at both ends with E2000 angle
connectors. Multimode optical fiber configured at
both ends with SMA connectors.
Cable sheath Oil-resistant polyurethane
Dimensions
- Diameter
- Length
< 8 mm,
up to 1 000 m
for lengths > 500 m, an external power supply
must be additionally ordered
for installations in Ex zones, non-intrinsically-safe
cables have to be spatially separated from
intrinsically-safe lines, lengths <600 m
Impact resistance 200 N/cm
Maximum tensile strength 500 N
Minimum bending radius 10 cm
Table 8- 11 Climatic conditions
Climatic conditions
Ambient temperature -40 ... +80 °C during operation
Rel. humidity < 95 % rel. humidity, above dew point