EasyManua.ls Logo

Siemens LDS 6 - Purging

Siemens LDS 6
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Data
8.3 Purging
LDS 6
88 Operating Instructions, 01/2009, A5E00295894-05
8.3 Purging
Conditions of purging media
As medium nitrogen is permissible for purging the sensor side. Nitrogen, steam, air and
gases which are not subject to the pressure equipment directive Cat. 2 are permissible as
purging gases for the process side.
Table 8- 12 Purging with air and nitrogen
Purging with instrument air, N2
Pressure at purging inlet 200 ... 800 kPa
Maximum overpressure in the sensor < 500 hPa
Quality
- Instrument air
- Nitrogen
free of oil and water
Purity better than 99.7 %. For oxygen
measurements, an O
2
content < 0.01 % is
recommended in the purging gas (optical path
length ≥ 1 m, min. 5 % oxygen in the process
gas)
Maximum flow rate 500 l/min
Dew point Benchmark: ≤10 °C,
condensation on the optics must be avoided
Table 8- 13 Blower purging
Blower purging
Maximum counter pressure 4 kPa
Maximum flow rate 350 l/min
Power consumption 370 W
Degree of protection (fan) IP54
Table 8- 14 Steam purging
Steam purging
Steam conditioning Overheated
Maximum temperature 240 °C
Minimum pressure > 400 kPa
Maximum pressure 1 600 kPa,
refers to a volume flow of approx. 1 100 l/min

Table of Contents

Other manuals for Siemens LDS 6

Related product manuals