Introduction
MAMMOMAT Inspiration 9
SPB7-330.640.01.01.02
For the qualification of the printer please follow the printer manufacturer’s
recommendations.
Important Notes
Before performing any quality control test, be sure to be familiar with the operating
instructions described in the MAMMOMAT Inspiration Operator Manual.
The enclosed test form tables can be used to document test parameters used and
results obtained during each test.
Recommended Corrective Action
Whenever there is a result from a test described in this manual that fails to reach the
action level stated, the problem must be detected and corrected before further
examinations are performed with the system. The result shall be reported to a service
engineer and proper corrective action shall be taken to address the problem.
If a result meets the action level, but a clear degradation is seen compared to the
result obtained during the latest test, the system may continue to be used, but the
degradation shall be reported to a service engineer, and it is recommended that
corrective action be taken within 30 days.
More details are given in association with each test.
Required Documents
• MAMMOMAT Inspiration Operator Manual
• Quality Control Manual of used viewing station (optional - for systems that
use softcopy review station)