EasyManua.ls Logo

Siemens MAMMOMAT Inspiration - AEC Test Performance Criteria

Siemens MAMMOMAT Inspiration
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Test 1: AEC Test
18 Quality Control Manual
SPB7-330.640.01.01.02
Performance Criteria and Corrective Action
The AEC security cut-off mechanism shall work properly in such a way that in cases
where no radiation reaches the detector no tube load beyond 5 mAs can be delivered.
The AEC segmentation should be able to select an AEC region that:
- completely remains inside the shadow of the PMMA plates, and
- whose borders are close to the ones of the PMMA.
The coefficient of variation for the mean pixel values should not exceed 5 % when
going through all PMMA thicknesses.
The maximum relative deviation for the mean pixel values should not exceed 10 %
when going through all PMMA thicknesses.